Purity is not content: HPLC, LC-MS and net peptide content
Three measurements that sound similar but answer different questions — and why a vial can be high purity and under-filled, or over-filled with low purity.
Three measurements that sound similar but answer different questions — and why a vial can be high purity and under-filled, or over-filled with low purity.

Three numbers on a peptide certificate get conflated more than any others: purity, identity, and net peptide content. They are measured by different methods, they answer different questions, and failing to keep them separate leads to exactly the kind of mistake a certificate is meant to prevent.
Purity is measured by reversed-phase HPLC. The chromatograph separates everything in the vial by how strongly each component sticks to the column, and a detector records a peak for each. Purity is the main peak's share of the total peak area — the percentage of everything detected that is the intended molecule. If the main peak is 99% of total area, purity is 99%. What purity does not tell you is how many milligrams of that 99%-pure material are in the vial.
Identity is measured by LC-MS. Liquid chromatography separates the components and mass spectrometry weighs them. The molecule has a theoretical mass, calculated from its amino-acid sequence; the mass spectrometer measures the observed mass. If observed and theoretical agree within tolerance, identity is confirmed — this is the molecule the label names. Identity is a yes/no on the molecule, not a quantity and not a purity.
Net peptide content is the actual mass of the peptide in the vial. It is the number of milligrams of the peptide itself, as distinct from water, salts, and other non-peptide material. A vial can hold 10 mg of total material that is 99% pure peptide — meaning about 9.9 mg of peptide — or it can hold 10 mg of total material that is 50% pure, meaning about 5 mg of peptide. Same vial weight, same purity number, very different amount of what you actually want.
The point is that these can move independently. A vial can be high purity and under-filled: the molecule is clean, but there just is not much of it. Or, as in the 2024 online semaglutide study, a vial can be over-filled with low purity: every sample held more semaglutide than its label claimed, yet purity was only about 7 to 14% against 99% advertised. "More than labeled" and "mostly not the molecule" were true at the same time.
This is why a certificate that reports only purity is incomplete. A buyer who sees "99% pure" and assumes the vial holds the labeled number of milligrams of that 99%-pure material is making two assumptions the number does not support. A complete COA reports identity (mass), purity (main-peak percent), and net content (milligrams), each with its method — so that the three questions are answered separately instead of being blurred into one.