ExplainerContamination and Safety2024

Heavy metals in synthesized products

Arsenic, cadmium, lead and mercury — where they come from, how they are measured, and the limits that apply.

Heavy metals in synthesized products
Key findings
  • Arsenic, cadmium, lead and mercury are elemental impurities from reagents, catalysts and equipment
  • Measured by ICP-MS per USP <233>, in ppm
  • Limits set by USP <232> and ICH Q3D — As 1.5, Cd 0.2, Pb 0.5, Hg 0.3 ppm
  • Heavy metals are invisible to a chromatography purity method, so they need their own line on a COA

Heavy metals are elemental impurities — arsenic, cadmium, lead, and mercury — and they are a concern in synthesized products because they can be introduced at several points in manufacturing and because they are toxic at low levels.

Where they come from. Peptides and small molecules are synthesized chemically, and that chemistry uses reagents, catalysts, and equipment. Metals can enter as catalysts, as contaminants in reagents, or from metal surfaces and vessels. Because they are elements, they do not break down — whatever arsenic or lead enters the process stays in the product unless it is removed, and removal is harder than avoiding introduction.

How they are measured. The method is ICP-MS — inductively coupled plasma mass spectrometry — run per USP <233>. The sample is ionized in a plasma, and the mass spectrometer counts the atoms of each element. ICP-MS is chosen because it can measure these elements at very low concentrations, in the parts-per-million range and below, which is where the relevant risk lives.

The limits. USP <232> and the international guideline ICH Q3D set permitted daily exposure limits for elemental impurities. The four that drive most peptide work are arsenic, cadmium, lead, and mercury, and the limits we test against — 1.5 ppm arsenic, 0.2 ppm cadmium, 0.5 ppm lead, and 0.3 ppm mercury — are drawn from those frameworks. A result is read against the permitted limit: below it passes, above it fails.

Why it matters. The risk from heavy metals is cumulative and not something a purity number captures. A product can be 99% pure by HPLC and still carry lead above its limit, because the lead is not a peak the chromatograph is looking at — it is an element, invisible to a chromatography purity method. That is why a heavy-metals line on a certificate of analysis is separate from the purity line, with its own method (ICP-MS), its own units (ppm), and its own specification (the USP <232>/ICH Q3D limits).

So when you read a COA, the heavy-metals result answers a question the chemistry results cannot: did the synthesis carry toxic elements into the vial? It is measured by a method built for trace elements, against limits built for safety, and reported in its own units so it cannot be confused with purity.

Cites: USP <232>, USP <233>, ICH Q3D.

Educational information only. Not medical advice. USA Prime Labs tests research materials for research use only; nothing here is intended for human or veterinary use. Talk to a licensed health care provider about any medical decision.
Related articles
USA Prime Labs
USAPRIME LABS
We test the research before you do

We test the research before you do. Identity, purity, content, endotoxin, metals, sterility — verified on one certificate you can check by QR.

All testing is performed on research materials for research use only. Not for human consumption. Results describe the sample as received and do not constitute medical, clinical, or safety advice.

© 2026 USA Prime Labs. Research use only — not for human consumption.Part of the USA Prime family.